Comparison of automated multiplexed bead-based ANA screening assay with ELISA for detecting five common anti-extractable nuclear antigens and anti-dsDNA in systemic rheumatic diseases

Clin Chim Acta. 2012 Jan 18;413(1-2):308-11. doi: 10.1016/j.cca.2011.10.017. Epub 2011 Oct 20.

Abstract

Background: A newly developed and totally automated Luminex-based assay, the BioPlex™ 2200 system, is able to detect various autoantibodies simultaneously from a single sample. We compared the BioPlex™ 2200 system with ELISA for the detection of six autoantibodies.

Methods: A total of 127 serum samples from the patients with systemic rheumatic diseases were collected and assayed with the BioPlex™ 2200 system (Bio-Rad, USA) and conventional ELISA (INOVA Diagnostics, USA) for 5 anti-extractable nuclear antigens. Additionally, relative sensitivity of the BioPlex™ 2200 system for detecting anti-dsDNA was evaluated with 79 specimens from SLE patients, which were positive for anti-dsDNA by ELISA.

Results: The concordance rates between ELISA and the BioPlex ranged from 88.1% for anti-RNP to 95.2% for anti-Scl-70, and the kappa coefficients between the results by the two assays were from 0.48 to 0.67. Among the 79 anti-dsDNA positive specimens by ELISA, seventy-eight (98.7%) showed positive results for anti-dsDNA by the BioPlex.

Conclusions: The BioPlex™ 2200 system showed comparable results with those by conventional ELISA for detecting autoantibodies, and this automated assay could measure multifarious autoantibodies concurrently in a single sample. It could be effectively used in clinical laboratories for screening autoimmune diseases.

Publication types

  • Comparative Study

MeSH terms

  • Antigens, Nuclear / analysis*
  • Automation*
  • DNA / immunology*
  • Enzyme-Linked Immunosorbent Assay / methods*
  • Humans
  • Microscopy, Fluorescence
  • Rheumatic Diseases / immunology*