TY - JOUR T1 - Successful Evaluation of Spinal Mobility Measurements With the Epionics SPINE Device in Patients With Axial Spondyloarthritis Compared to Controls JF - The Journal of Rheumatology JO - J Rheumatol SP - 44 LP - 52 DO - 10.3899/jrheum.201470 VL - 49 IS - 1 AU - David Kiefer AU - Xenofon Baraliakos AU - Daniela Adolf AU - Varvara Chatzistefanidi AU - Ilka Schwarze AU - Uwe Lange AU - Jan Brandt-Jürgens AU - Edgar Stemmler AU - Sabine Sartingen AU - Jürgen Braun Y1 - 2022/01/01 UR - http://www.jrheum.org/content/49/1/44.abstract N2 - Objective Epionics SPINE (ES), a novel device that measures spinal movements using electronic sensors including range of motion (RoM) and speed (range of kinematics [RoK]), has already been validated in patients with mechanical back pain and healthy individuals. This study aimed to evaluate ES for quantification of spinal mobility in patients with axial spondyloarthritis (axSpA).Methods A total of 153 individuals, 39 female and 114 male, were examined including 134 patients with axSpA, of whom 40 had nonradiographic (nr)-axSpA, 94 had radiographic (r)-axSpA; 19 were healthy controls (HCs). The results were compared using mean ES scores and modeling was performed using multivariable logistic regression models resulting in good validity and high discriminative power.Results ES measurements showed meaningful differences between patients with axSpA and HCs (all P < 0.001), as well as between r- and nr-axSpA (P < 0.01). In patients with axSpA, a negative correlation between ES and Bath Ankylosing Spondylitis Metrology Index values was found: –0.76 ≤ r ≤ –0.52 (P < 0.05). Bath Ankylosing Spondylitis Functional Index scores showed a similar trend (r > –0.39). Patients with r-axSpA had a more limited and slower spinal mobility than those with nr-axSpA. Other patient-reported outcomes almost did not correlate.Conclusion This study shows that the ES is an objective performance measure and a valid tool to assess spinal mobility in axSpA, also based on the Outcomes Measures in Rheumatology (OMERACT) criteria. RoK and RoM scores provide additional information on physical function of patients with axSpA. ER -