Seronegative spondyloarthropathy (SpA) is a chronic inflammatory disease that includes a heterogeneous group of diseases such as psoriatic arthritis (PsA), ankylosing spondylitis (AS), reactive arthritis, arthritis associated with inflammatory bowel disease, and undifferentiated SpA. Enthesis and tendon involvement are 2 important aspects of SpA1.
There is strong evidence that demonstrates the role of ultrasound (US) with power Doppler (PD) technique and its higher sensitivity, compared to the clinical examination, in the detection of signs of entheseal involvement in both late and early stages2. One of its main limits is operator dependency. A minimal change in the transducer position or an erroneous setting can provide completely different information, easily leading to misinterpretations.
The importance of the patient’s position during the US assessment has not been well studied3,4. We investigated the effect of this phenomenon at different enthesis levels in 4 patients with SpA. US images were obtained with a MyLab 70 XVG US system (Esaote Biomedica, Genoa, Italy) with a 6–18 MHz linear transducer B-mode frequency of 18 MHz and a Doppler frequency of 9.1 MHz. To our knowledge no similar studies have been conducted.
The first case was a 29-year-old man with entheropathic-related SpA. He reported 2 weeks of anterior knee pain. Figure 1 shows the US findings at the quadriceps tendon insertion, taken with the patient in neutral position and 30° of knee flexion. The main change is characterized by completed is appearance of the intratendinous PD signal.
The second case was a 32-year-old man with PsA who had knee pain for 1 month. US examination performed at the proximal insertion of the patellar tendon, adopting both positions (neutral and flexed), showed a marked decrease of PD signal after the knee flexion (Figure 2).
The third case was a 16-year-old girl with a diagnosis of undifferentiated SpA and 3 months of knee pain. A dramatic change in terms of PD signal was seen when the patient adopted the knee flexion position (Figure 3).
The fourth case was a 36-year-old man with a diagnosis of AS. US examination performed at the insertion of the Achilles tendon in neutral position showed an enthesopathy associated with intense PD signal within both the tendon and the retrocalcaneal bursa. This PD finding disappeared completely after foot dorsiflexion (Figure 4).
Guidelines for the use of grayscale US in the musculoskeletal system have been developed by the European League Against Rheumatism work group5. However, no guidelines have been formulated for Doppler assessment, which appears particularly sensitive to several factors (such as setting, artifacts, and transducer pressure)6. In specific cases, the flexion of both knee and ankle could determine an increase of the intratendinous tension, which facilitates the collapse of the microvessels. Since PD visualization at the entheseal level is an important indicator of disease activity in SpA, we suggest further consideration be given to patient position during the US examination in order to determine accurately the entheseal inflammatory involvement.
Our report provides visual evidence that patient position plays an important role in the PD assessment of entheses. Further studies of validity are required to confirm our findings.