To the Editor:
After reading the interesting review by Gutierrez, et al1, entitled “Ultrasound in the assessment of interstitial lung disease in systemic sclerosis. A systematic literature review by the OMERACT Ultrasound Group,” we must take issue with some of the statements published.
The authors state that “B-lines consist of ‘comet tails’… generated by the reflection of the [lung] US beam from thickened subpleural interlobar septa.” Actually, the generation of ultrasound (US) artifacts mainly depends on the high difference in acoustic impedance that the US beam encounters when it crosses surfaces with a different density. US scanner machines are calibrated at a constant sound speed of about 1500 m/s, but propagation velocity in air is only 330 m/s (in the lung, slightly increased to 440 m/s owing to presence of parenchyma): for this reason more than 96% of the US beam is reflected at tissue–chest wall/air-lung interface, resulting in hyperechoic pleural line and US artifacts2. In particular, B-lines, appearing as parallel hyperechogenic striae extended for the entire US image, result from different acoustic impedance between liquid film trapped in gas microbubbles and/or fibrous edema areas crossed by the US beam. That is why B-lines are also found in the bowel loops3 and in the residual cavity of the postpneumonectomy space4. On the other hand, “comet tails,” running out in a few mm of depth in a triangular shape, derive from solid structures such as crystals (e.g., cholesterin, calcium) or pleural thickenings. Therefore, “B-lines” and “comet tails” are different in shape and content.
Number of B-lines may increase in all those pathological pleuropulmonary conditions in which the proportion of air/liquid film changed: heart failure, adult respiratory distress syndrome, pneumothorax and hydropneumothorax, fibrosis, emphysema, exacerbations of chronic obstructive pulmonary diseases, nephrotic syndrome, and lymphagitis5,6,7. This type of US artifact is not a pathognomonic index of fibrosis: for this reason it is not possible to assess the presence of an initial interstitial lung disease in systemic sclerosis (SSc) only by detecting and/or counting B-lines. Recently we studied patients with a histological diagnosis of fibrosis, performing both transthoracic US (TUS) and an intraoperative ultrasound examination during video-assisted thoracic surgery (VATS). Generation of B-lines artifacts did not occur in our intraoperative examination in VATS-US, despite the presence of B-lines in TUS. Indeed, in areas with pulmonary fibrosis (PF), only a thicker hyperechoic line compared to the nonfibrotic lung was visible, in the absence of any other artifact8,9. Unlike TUS, the VATS-US approach is not limited by differences in acoustic impedance, because the probe is directly in contact with the lung (Figure 1). This, therefore, confirms how the high difference in acoustic impedance between chest wall and air influences the visualization of the pleurae and the lungs during TUS, and consequently generates artifacts such as B-lines.
A. Transthoracic ultrasound (TUS) with convex probe (5 MHz) showing irregular thickening (> 3 mm) of the hyperechoic pleural line (black arrow) and B-lines artifacts (white arrows) below it. B. Computed tomography axial scan of the same image visualized at TUS (black box). C. VAT intraoperatory US in the same patient with fibrosis. D. Intraoperatory US image with linear probe (10 MHz) showing increased thickness of pleural surface (white arrow) in the absence of B-lines below. VAT: video-assisted thoracic.
The number of B-lines also depends on the probe used, the degree of total gain compensation, and tissue harmonic and electronic beam focusing. High frequency probes (linear probe, 8–12.5 MHz) reduce the number of artifacts; sectorial probes (2–3.5 MHz) create a folding in the US beam that increases the number of artifacts. The types of probes used in the studies considered by the review authors vary; they are an important cause of bias in detection/count of B-lines.
Ultimately, TUS can at best explore no more than 70% of the most superficial lung pleura facing the chest surface, where it is not obscured by bones or other structures, such as the parietal mediastinal pleura. Although this limits the effectiveness of the TUS versus high-resolution computed tomography (HRCT), PF in SSc tends to originate at the posterior and basal pulmonary fields, where TUS exploration is not critically impaired. Pleural line thickness assessed by TUS displays good concordance with HRCT patterns, indicating lung fibrosis and changes attributable to its progression10. In this context TUS may become a useful complementary tool in well-trained hands for indicating timely HRCT assessment of PF in SSc in early diagnosis and followup.