TY - JOUR T1 - Physician control of needle and syringe during aspiration-injection procedures with the new reciprocating syringe. JF - The Journal of Rheumatology JO - J Rheumatol SP - 771 LP - 778 VL - 33 IS - 4 AU - Wilmer Sibbitt, Jr AU - Randy R Sibbitt AU - Adrian A Michael AU - Druce I Fu AU - Hilda T Draeger AU - Jon M Twining AU - Arthur D Bankhurst Y1 - 2006/04/01 UR - http://www.jrheum.org/content/33/4/771.abstract N2 - OBJECTIVE: To evaluate physician control of needle and syringe during aspiration-injection syringe procedures by comparing the new reciprocating procedure syringe to a traditional conventional syringe. METHODS: Twenty-six physicians were tested for their individual ability to control the reciprocating and conventional syringes in typical aspiration-injection procedures using a novel quantitative needle-based displacement procedure model. Subsequently, the physicians performed 48 clinical aspiration-injection (arthrocentesis) procedures on 32 subjects randomized to the reciprocating or conventional syringes. Clinical outcomes included procedure time, patient pain, and operator satisfaction. Multivariate modeling methods were used to determine the experimental variables in the syringe control model most predictive of clinical outcome measures. RESULTS: In the model system, the reciprocating syringe significantly improved physician control of the syringe and needle, with a 66% reduction in unintended forward penetration (p < 0.001) and a 68% reduction in unintended retraction (p < 0.001). In clinical arthrocentesis, improvements were also noted: 30% reduction in procedure time (p < 0.03), 57% reduction in patient pain (p < 0.001), and a 79% increase in physician satisfaction (p < 0.001). The variables in the experimental system--unintended forward penetration, unintended retraction, and operator satisfaction--independently predicted the outcomes of procedure time, patient pain, and physician satisfaction in the clinical study (p < or = 0.001). CONCLUSION: The reciprocating syringe reduces procedure time and patient pain and improves operator satisfaction with the procedure syringe. The reciprocating syringe improves physician performance in both the validated quantitative needle-based displacement model and in real aspiration-injection syringe procedures, including arthrocentesis. ER -